Leem Lab · UT Dallas · MoS₂ Raman

Raman Si Calibrator

Drop raw Raman TXT exports. Every spectrum is shifted so its Si peak sits at the reference, then E¹₂g, A₁g and the peak difference are measured. You get a formatted Excel workbook and a ZIP of plots.

Processed in your browser.

Raw data

TXT exports straight from the spectrometer software. One or many spectra per file, any number of files, 1800 or 600 gr/mm.

    Settings

    Results

    Click a row to see that spectrum's plot. Tick the boxes to stack several spectra for comparison.

    Files
    0
    Spectra
    0
    With MoS₂
    0
    Median Δ (cm⁻¹)
    –
    PickFile · #Δ (cm⁻¹)Si rawShiftE¹₂gA₁g
    Δ 18 to <20 monolayer-like 20 to <23 bilayer-like 23 to 30 few-layer / thick

    How it works

    Each spectrum is shifted so its own Si peak sits at the reference, then E¹₂g and A₁g are located between data points by a Gaussian fit through each peak top. Their difference sorts every spot into a layer class.

    Si calibration

    The Si peak (490 to 555 cm⁻¹) gets a straight baseline, then a Gaussian fit through its top gives the position between data points. A flat, saturated top is located from the midpoint of its two sides at 75% height. The whole spectrum is shifted so Si lands on the reference.

    MoS₂ peaks

    On the shifted axis, E¹₂g (372 to 394) and A₁g (396 to 418) are found the same way, with widths read at half height. The peak difference is A₁g minus E¹₂g; the Si shift moves both peaks equally, so it never changes the difference.

    Accuracy

    A spectrum counts as MoS₂ when A₁g is more than 6 times the noise. On coarse 600 gr/mm scans the difference is good to about ±0.5 cm⁻¹. Values within 0.3 of a class boundary are borderline.